Kern DIS Moisture Analysers


Kern DIS Moisture Analyser

Kern DIS moisture analysers use force-compensation weighing and halogen drying for laboratory material tests. The DIS 100-3 provides 120 g capacity with 0.001 g weighing readability; the DIS 50-4 provides 50 g capacity with 0.0001 g readability. Moisture readability is 0.001% and 0.0001% respectively.

Both models operate from 30 to 175 °C and use a 95 mm sample pan. The graphic LCD displays the drying curve so operators can follow changes during a test. Standard, rapid, gentle and staged drying profiles, stored methods and programmable sample descriptions support the development and repetition of material-specific procedures.

Model: DIS 50-4

Model

Calibration Certificate: Yes

Calibration Certificate

Sale price£1,883.00

Tax excluded. Shipping calculated at checkout

Estimated delivery: Approx. 7 working days.

Includes 2 additional working days for calibration.

Learn more

Product specification

Compare model specifications.

Model DIS 100-3 DIS 50-4
Capacity 120 g 50 g
Readability 1 mg 0.1 mg
Moisture Readability 0.001% 0.0001%
Repeatability (Weighing) 0.002 g 0.0002 g
Legal For Trade No No
Platform Size 95 mm 95 mm
Units g, % moisture, % dry mass, %M ATRO, g/l, % R (ratio), %D ATRO g, % moisture, % dry mass, %M ATRO, g/l, % R (ratio), %D ATRO

Learn more

Applications

Fine-Resolution Material Testing

Use the DIS 50-4 where a checked method requires 0.0001 g weighing readability and small moisture display increments.

Higher-Capacity Laboratory Samples

Select the DIS 100-3 for prepared samples requiring up to 120 g weighing capacity.

Drying Method Assessment

Review drying curves when comparing heating settings and deciding on an appropriate test end point.

Repeated Material Quality Checks

Recall stored procedures and identify samples when testing recurring materials or production batches.

Features

Two Force-Compensation Models

Select 120 g capacity with 1 mg weighing readability or 50 g capacity with 0.1 mg readability, according to the test requirement.

Drying Curve Display

View the drying curve on the graphic LCD to follow sample behaviour and assess the progress of a test.

Adjustable Heating Profiles

Standard, rapid, gentle and staged profiles provide control over how samples are heated between 30 and 175 °C.

Method and Sample Management

Stored procedures, sample descriptions and USB data transfer help organise recurring tests and their results.

FAQs

Need more help? Contact our sales team.