A 100 g test weight is not necessarily accurate enough for every balance that can weigh 100 g. The nominal mass tells you the load; the accuracy class tells you the permitted mass error. The physical design, certificate and way you handle the weight also matter.
This guide explains what to compare before buying, with examples from our Kern and Adam Equipment test weights.
Start with the job you need the weight to do
For a daily check at one working load, an individual weight may be enough. To check several loads, a set can be more useful. If you are following the balance’s external adjustment routine, check its manual for the specified mass and class before ordering.
Routine checking, calibration and adjustment are different activities. A check monitors performance at selected loads. Calibration establishes the relationship between the instrument indication and reference values, with uncertainty. Adjustment changes the instrument’s response. Buying a weight does not, by itself, complete any of these procedures.
What do M1, M2, F1, F2, E1 and E2 mean?
These are accuracy classes in OIML R111, which sets metrological and physical requirements for weights. At the same nominal mass, E1 has tighter tolerances than E2, followed by F1, F2, M1 and M2. The class is not one fixed percentage accuracy.
| Class | 100 g mass tolerance | Browse the range |
|---|---|---|
| M1 | ±5 mg | M1 weights and sets |
| M2 | ±16 mg | M2 weights and sets |
| F1 | ±0.5 mg | F1 weights and sets |
| F2 | ±1.6 mg | F2 weights and sets |
| E1 | ±0.05 mg | E1 weights and sets |
| E2 | ±0.16 mg | E2 weights and sets |
These figures apply to 100 g only. They are class limits, not the measured error or calibration uncertainty of an individual weight. Conformity assessment also takes account of uncertainty. R111 includes M3 and intermediate classes for specified applications; E3 and F3 are not R111 classes.
Which class should you buy?
Choose from your acceptance limits, rather than from capacity alone. A production scale and an analytical balance may both accept a 100 g load but need very different reference weights.
M1 and M2: compare the permitted error first
M1 and M2 are worth comparing for weighing checks where wider reference tolerances are acceptable. M2 has wider tolerances than M1 at the same nominal mass. It can be a suitable option when your check limits allow it, but a lower purchase price is of little benefit if the reference is too uncertain for the job.
For a repeated check using one stainless steel mass, compare our Kern M1 individual weights. For heavier loads, the Kern M2 cast iron blocks provide a different physical format; check platform space and handling arrangements.
F1 and F2: tighter reference tolerances
F1 and F2 give you tighter class tolerances than M1. Compare them when the allowed weighing error calls for a finer reference, rather than treating either as suitable for every precision balance.
The Kern F1 individual range is one option for single test points. Full sets and Safety Sets cover different checking routines; choose the masses you actually need, not simply the widest available range.
E1 and E2: laboratory and reference-mass work
E1 and E2 deserve consideration where your laboratory procedure requires tighter reference tolerances. They also bring greater demands on documentation, handling and conditions. More expensive weights will not solve problems caused by draughts, vibration, contamination or an unsuitable balance.
Compare Adam E2 stainless steel sets for several laboratory test loads, or Kern E1 sets where the required uncertainty justifies that class.
Why readability is not enough to select a weight
Readability is the smallest displayed increment. It does not tell you the error allowed in your process, or the uncertainty needed for your reference. Start with the manufacturer’s instructions and your written checking procedure.
For example, if a hypothetical 100 g check permits an instrument error of only ±2 mg, an M1 reference with a class limit of ±5 mg is too coarse to use on nominal value alone. F2 has a smaller limit of ±1.6 mg, but that does not automatically make it suitable: you still need to account for certificate results, uncertainty and your decision rule. F1 may offer more room in that assessment.
This example illustrates the comparison, not a universal selection rule. If your procedure specifies a particular uncertainty ratio or correction method, follow it. Tell us your permitted error as well as the instrument’s capacity and readability.
Individual weights, full sets or Safety Sets?
- Individual weights: useful for a repeated check at one load, replacing a missing mass or extending a set you already hold.
- Full sets: provide several masses and combinations. Check repeated values as well as the smallest and largest weight.
- Safety Sets: contain two selected reference weights for defined smaller and larger test loads. They do not supply every intermediate mass.
A set described as 1 mg–200 g has 200 g as its largest individual weight. That is not the combined mass of the set. The two Adam E2 set options, for example, both finish at 200 g, but one starts at 1 mg and the other at 1 g.
Choose a physical design you can use and store properly
Knob and compact weights suit different pans and handling methods. Small wire or platelet weights need appropriate tools: compare the Kern E2 wire weights and E2 platelet weights if you need individual milligram references.
Larger block and check weights need enough platform space and a planned handling method. Stainless steel and lacquered cast iron have different care requirements. Inspect cast iron finishes for damage or corrosion, and keep all reference surfaces free from contamination.
Check storage before ordering. Sets include the case specified for the model. Adam individual stainless steel weights include plastic cases; Kern matching plastic boxes are optional where shown. Do not assume that every large weight includes a case.
Weight class, calibration and accreditation are separate questions
A class describes requirements for the weight. Calibration provides results for the individual item. Accreditation assesses the laboratory’s competence for a defined scope of work, generally against ISO/IEC 17025; it is not another weight class.
OIML R111 requires class E weights to be accompanied by a calibration certificate. For E1 or E2 work, confirm the appropriate certificate and reported data before ordering. Do not assume that selecting a weight without the optional certificate provides the documentation needed for its intended use.
Check that the certificate identifies the weight and reports the mass result, uncertainty and other information needed by your procedure. “Traceable” does not automatically mean “UKAS-accredited”. If accredited calibration is required, check the laboratory’s current scope and measurement capability for your masses and required uncertainty.
On our listings, the available choices depend on the manufacturer and class:
- Kern: optional DAkkS calibration is offered on the relevant listings. DAkkS is Germany’s accreditation body; it is distinct from UKAS. Confirm acceptance if your contract names a specific accreditation requirement.
- Adam M1 and F1: Traceable or UKAS options, with Traceable referring to Adam’s in-house calibration statement.
- Adam E2: a UKAS certificate option.
- Adam E1: calibration by quotation; the displayed weight prices exclude an optional certificate.
Our guide to instrument calibration explains the difference between results, adjustment and conformity decisions in more detail.
Newton weights: for force rather than mass checks
A newton is a unit of force; a gram or kilogram is a unit of mass. The force produced by a suspended mass depends on gravity. Do not assume that a 1 kg mass provides exactly 10 N.
Our Newton weights are specified for reference-force work and adjusted for a stated place of use. Choose hook weights for compatible suspended arrangements, or Type A slotted weights with the matching Type A supporting bar. Include the bar’s own load in the total applied force and check compatibility before mixing components.
Looking after your reference weights
Use suitable gloves or forceps, keep cases clean and allow weights to reach the weighing-room temperature before precision checks. Do not polish, engrave or alter a reference weight yourself. A surface that looks clean can still carry deposits that affect mass.
Set recalibration intervals through your quality system, using the history, frequency of use, handling and consequences of an incorrect result. Review the interval or seek advice after damage, corrosion, contamination or unexpected check results. An annual interval is not automatically right for every weight.
What to tell us before you order
- Instrument: manufacturer, model, capacity and readability.
- Test loads: the loads you actually use, plus any mass specified for external adjustment.
- Acceptance limits: the permitted error and any uncertainty or ratio requirement.
- Documentation: the certificate and accreditation requirements in your procedure or contract.
- Practical details: pan or platform space, handling and storage. For Newton weights, include the place of use and fixture details.
Browse all test weights, or contact GNW for help choosing. Those details let us compare a suitable reference with your actual requirements.